Stimulací indukovaná kardiomyopatie
| Title in English | Pacing-induced cardiomyopathy |
|---|---|
| Authors | |
| Year of publication | 2024 |
| Type | Peer-reviewed scientific article |
| Magazine / Source | COR ET VASA |
| MU Faculty or unit | |
| Citation | |
| web | https://e-coretvasa.cz/pdfs/cor/2024/06/08.pdf |
| Doi | https://doi.org/10.33678/cor.2024.054 |
| Keywords | cardiac pacing; cardiac resynchronization therapy; conduction system pacing; Pacing induced cardiomyopathy |
| Description | Pacing-induced cardiomyopathy (PICM) does not have an internationally accepted definition. In general, the term refers to the heart failure with a left ventricular EF below 50% and its reduction ? 10% after implantation of a pacemaker without known cause. It is a complication of a conventional cardiac pacing, which is associated with a higher cardiovascular and all-cause mortality and more frequent hospital admissions. The pacing burden that increases the risk of PICM is > 20%. On the other hand, it cannot be said that a lower pacing burden is safe or that a higher pacing burden is clearly harmful. It is necessary to identify patients with a higher risk of developing PICM before the implantation of a pacemaker and offer other (for now) alternative methods of pacing. In patients with a conventional cardiac pacing, it is necessary to optimize the programming to suppress unwanted ventricular pacing with the aim of reducing the risk of PICM as much as possible. In the case of an already developed PICM, an upgrade to a CRT or a CSP should be considered. If indicated it is recommended to perform the upgrade without unnecessary delay after the recognition of the PICM. |