A new liquid chromatography-tandem mass spectrometry method using atmospheric pressure photo ionization for the simultaneous determination of azaarenes and azaarones in Dutch river sediments

Logo poskytovatele

Varování

Publikace nespadá pod Lékařskou fakultu, ale pod Přírodovědeckou fakultu. Oficiální stránka publikace je na webu muni.cz.
Autoři

BRULÍK Jan ŠIMEK Zdeněk DE VOOGT Pim

Rok publikování 2013
Druh Článek v odborném periodiku
Časopis / Zdroj Journal of Chromatography A
Fakulta / Pracoviště MU

Přírodovědecká fakulta

Citace
Doi http://dx.doi.org/10.1016/j.chroma.2013.03.079
Obor Znečištění a kontrola vody
Klíčová slova Azaarenes; Azaarones; PAH; APPI; LC-MS/MS; Sedimets
Popis A new method for the analysis of azaarenes and their degradation products (azaarones) was developed, optimized and validated using liquid chromatography coupled with atmospheric pressure photo ionization tandem mass spectrometric detection (LC-APPI/MS/MS). Seventeen compounds including 4 PAHs (naphthalene, anthracene, phenanthrene, benz[a]anthracene), 7 azaarenes (quinoline, acridine, phenanthridine, 5,6-benzoquinoline and 7,8-benzoquinoline, benzo[a]acridine, benzo[c]acridine), and 6 azaarones (2-OH-quinoline, 4-OH-quinoline, 5-OH-quinoline, 6-OH-quinoline, 9(10H)-acridone, 6(5H)phenanthridinone) were analyzed in sediment samples from Dutch rivers. All compounds were analyzed simultaneously in multi reaction monitoring (MRM) mode. Soxhlet extraction was used for the extraction of analytes from sediments. The limits of quantification of azaarenes and azaarones varied from 0.21 to 1.12 mu g/l and from 0.23 to 1.58 mu g/l, respectively. The limits of quantification for PAHs varied from 32 to 769 mu g/l. Matrix-independent recoveries of sediment samples were in the range 85-110%; matrix-dependent recoveries were in the range 73-148%, respectively. The method was tested on real sediment samples and the results were compared with a previous study in which GC/MS/MS was used for the simultaneous measurement of azaarenes and azaarones. 4-, 5- and 6-OH-quinolines and naphthalene, anthracene and phenanthrene were not present or below detection limits in some samples. All other analytes were present in samples in the concentration range 0.2-1200 ng/g (dw). To our knowledge, this is the first report showing the possibility of measurement non-polar polyaromatic hydrocarbons together with polar azaarenes and their degradation products azaarones simultaneously with sufficient sensitivity and accuracy using LC/MS/MS.
Související projekty:

Používáte starou verzi internetového prohlížeče. Doporučujeme aktualizovat Váš prohlížeč na nejnovější verzi.

Další info