Automated seizure detection using wearable devices: A clinical practice guideline of the International League Against Epilepsy and the International Federation of Clinical Neurophysiology

Autoři

BENICKY Sandor WIEBE Samuel JEPPESEN Jesper TATUM William O. BRÁZDIL Milan WANG Yuping HERMAN Susan T. RYVLIN Philippe

Rok publikování 2021
Druh Článek v odborném periodiku
Časopis / Zdroj Clinical Neurophysiology
Fakulta / Pracoviště MU

Lékařská fakulta

Citace
www https://www.sciencedirect.com/science/article/pii/S1388245720306143
Doi http://dx.doi.org/10.1016/j.clinph.2020.12.009
Klíčová slova Algorithms; Automated detection; Epilepsy; Seizure detection; Wearable devices
Popis The objective of this clinical practice guideline (CPG) is to provide recommendations for healthcare personnel working with patients with epilepsy, on the use of wearable devices for automated seizure detection in patients with epilepsy, in outpatient, ambulatory settings. The Working Group of the International League Against Epilepsy and the International Federation of Clinical Neurophysiology developed the CPG according to the methodology proposed by the ILAE Epilepsy Guidelines Working Group. We reviewed the published evidence using The Preferred Reporting Items for Systematic Review and Meta-Analysis (PRISMA) statement and evaluated the evidence and formulated the recommendations following the Grading of Recommendations Assessment, Development and Evaluation (GRADE) system. We found high level of evidence for the accuracy of automated detection of generalized tonic-clonic seizures (GTCS) and focal-to-bilateral tonic-clonic seizures (FBTCS) and recommend use of wearable automated seizure detection devices for selected patients when accurate detection of GTCS and FBTCS is recommended as a clinical adjunct. We also found moderate level of evidence for seizure types without GTCs or FBTCs. However, it was uncertain whether the detected alarms resulted in meaningful clinical outcomes for the patients. We recommend using clinically validated devices for automated detection of GTCS and FBTCS, especially in unsupervised patients, where alarms can result in rapid intervention (weak/conditional recommendation). At present, we do not recommend clinical use of the currently available devices for other seizure-types (weak/conditional recommendation). Further research and development are needed to improve the performance of automated seizure detection and to document their accuracy and clinical utility. (c) 2020 International Federation of Clinical Neurophysiology, Inc. and International League Against Epilepsia. Published by Elsevier B.V. All rights reserved.

Používáte starou verzi internetového prohlížeče. Doporučujeme aktualizovat Váš prohlížeč na nejnovější verzi.

Další info